NOTE: See "Daily Activities" on the Home page for daily maintenance
routines.
Luminex recommends that you verify daily and calibrate weekly. You should also verify and
calibrate if any of the following occurs:
The delta calibration temperature exceeds ± 5° C.
You move the instrument.
You experience sample acquisition problems.
You perform maintenance on the instrument, for example, replacing a hardware
component.
Available system initialization procedures are as follows:
Fluidics preparation, calibration, performance verification
Fluidics preparation, performance verification
Fluidics preparation
Defining the System Initialization Routine
NOTE: Luminex recommends daily performance verification and weekly
calibration of the MAGPIX system. You can set up the system
initialization routine to include calibration and verification on the
Admin page, System Setup tab, Maintenance Options section.
1. Open the Admin page.
2. Click System Setup.
Admin Page
127